Response to Low ppm Readings on a Hydrogen Peroxide Monitor

Many users are concerned about exposure to hydrogen peroxide vapor from their sterilizers and so have invested in continuous monitors for hydrogen peroxide. These monitors provide a real time reading of the hydrogen peroxide concentration in ppm but now the user needs to interpret these numbers and respond appropriately.

If there is a large leak of hydrogen peroxide vapor and the hydrogen peroxide monitor is in high alarm, the response is simple – clear everyone out of the immediate area until the ventilation reduces the concentration to safe levels (as determined by the monitor). Fortunately massive leaks rarely occur, but it is not uncommon for people to see low ppm readings on their hydrogen peroxide monitors. The question then arises as to what constitutes over exposure to hydrogen peroxide vapor and what actions should be taken.

Some people rely only on the OSHA permissible exposure limit (1 ppm, calculated as a time weighted average (TWA) over 8 hours). The argument goes that so long as the OSHA PEL is not exceeded all is good in the world of hydrogen peroxide exposure. Sometimes this calculation is tempered by saying that the hydrogen peroxide exposure should not go over the NIOSH IDLH (75 ppm for hydrogen peroxide).

Following this logic, it would be OK for someone to be exposed to 50 ppm hydrogen peroxide continuously so long as it does not exceed the PEL, i.e. < 480 minutes/50 ppm = < 9.6 minutes, and 25 ppm for < 19.2 minutes etc. are OK. I would not want to be the person exposed and certainly not on a regular basis. These numbers may seem high but we have seen hydrogen peroxide vapor in the 30 to 40 ppm range being emitted from one model of hydrogen peroxide sterilizer whenever the door was opened after completion of a cycle. Hydrogen peroxide has essentially no odor and so apart from some a monitor or other method to detect hydrogen peroxide there is no way to know if hydrogen peroxide vapor is present.

Probably the most respected industrial hygiene organization in the world is the ACGIH and the ACGIH has issued a threshold limit value for hydrogen peroxide of 1 ppm calculated as an 8 hour TWA. The similarity with the OSHA PEL is not coincidental since the OSHA PELs were initially derived from the ACGIH TLVs in 1972 and neither the ACGIH TLV nor OSHA PEL have revised the exposure limit for hydrogen peroxide since then.

For compounds with no short term exposure limits the ACGIH recommends the following: Excursions in worker exposure levels may exceed 3 times the TLV-TWA for no more than a total of 30 minutes during a work day, and under no circumstances should they exceed 5 times the TLV-TWA, provided the TLV-TWA is not exceeded. [2008 TLVs and BEIs based on the Documentation of the threshold Limit Values for Chemical Substances and Physical Agents and Biological Exposure Indices, p 5.]

Applying this guidance to hydrogen peroxide, worker exposure may exceed 3 ppm for no more than 30 minutes during the work day and under no circumstances should they exceed 5 ppm. These levels are thus similar to a short term exposure limit (STEL) and ceiling limit respectively and are consistent with the STELs for hydrogen peroxide promulgated by several governmental occupational safety agencies. Washington state for example has a STEL of 3 ppm (15 min TWA) and the United Kingdom and some other European countries have a 2 ppm STEL. [Ref EH40, 2005]. While there is no OSHA STEL for hydrogen peroxide, this ACGIH guidance represents best practice when using hydrogen peroxide.

If the monitor always reads less than 1 ppm then the user need have no immediate concerns about exposure to hydrogen peroxide, the 8 hour TWA will be less than the OSHA PEL. However, we have found on many occasions that the readings often start our small but over time they increase. Thus if the readings increase over successive cycles of the sterilizer and become significantly higher than previously seen, even if still within safe limits, then these numbers can serve as an indicator that the sterilizer should be serviced soon.

If the readings occasionally rise between 1 to 5 ppm, as may some times occur when the sterilizer door is opened, then the user should step away from sterilizer and return once the readings have fallen to safe levels (< 1 ppm). If this ‘puff’ of hydrogen peroxide is new to the operation of the sterilizer, then again, it is time for maintenance.

If the hydrogen peroxide monitor goes above 5 ppm, then this concentration poses a potential hazard to employees and the problem should be rectified and the sterilizer manufacturer should be asked to correct the problem. If the manufacturer says that the sterilizer or other equipment is operating normally, then other measures such as engineering controls, modified work practices etc. should be employed to ensure workers are not exposed to hydrogen peroxide above 5 ppm.

Friday, July 6, 2012

Exposure Badges Do Not Provide Adequate Protection Against Toxic Gases

Toxic gases are found in many workplaces across many industries ranging from carbon monoxide in coal mines, hydrogen sulfide in waste water treatment and chlorine in paper mills. Typically, the workplace is designed to minimize worker exposure to these gases through work practices, process design, PPE & engineering controls etc. However, it is widely recognized that despite the best efforts of the design engineers, exposures to toxic gases can still occur and so most facilities using toxic gases employ fixed and/or portable instruments that continuously monitor the atmosphere and provide a warning to people working there in the event of a leak.

Employers provide continuous monitors because in part because they want to protect their workers and in part because they are required to provide a safe work environment. The Occupation Health and Safety Act of 1970 requires an employer to "furnish to each of his employees employment and a place of employment which are free from recognized hazards that are causing or are likely to cause death or serious physical harm to his employees". This statute provides a general legal duty, and OSHA's standards, which provide more specific legal requirements, such as maximum permissible exposure limits (PELs) for toxic gases. However, OSHA recognizes that each workplace is different and so OSHA leaves determining the best means to achieve these levels to the employer.

Prior to the development of continuous monitors, several chemical methods were developed for determining workplace exposures of toxic gases. Three common methods are:

1) Drawing a known volume of air through an impinger (bubbler) containing a reagent that reacts with the target gas, and the resulting product is then analyzed by laboratory.

(2) Gas exposure badges, in which the target gas reacts with a reagent impregnated on the surface of the badge and the product is again analyzed by a laboratory and

(3) Gas detection tubes in which the target reacts reacts with a reagent on a solid support giving a color change such that the extent of the color change along a graduated scale provides an indication of the gas concentration.

For routine monitoring, these methods have largely been replaced by continuous monitors. Impingers are still used today since many standard test methods call for them, and impingers and badges have some value for those 'exotic gases' for which continuous monitors are not available. The primary drawback of impingers and badges is that they do not provide any warning of what the worker is currently being exposed to, but only report what he or she has already been exposed to. For someone inhaling a toxic gas, now is important, two weeks time, or however long it takes for the lab reports to come back is not adequate. If a person protected by a continuous monitor is exposed to a high concentration of the toxic gas or vapor, a continuous monitor will alarm before the gas concentration reaches dangerous levels and so prevent that person from being exposed. Prevention is always better than an apology!

For this reason, one almost never sees exposure badges being used in general industry. They don't use exposure badges in water treatment plants, or chemical plants or other industries for gases for which continuous monitors are available. There is however one exception, the health care industry. The health care industry uses many toxic chemicals such as drugs for angioplasty, anesthetic gases and sterilization/high level disinfection. The high level disinfectants in particular included gases and vapors such as ethylene oxide (OSHA PEL 1 ppm, NIOSH IDLH 800 ppm) and hydrogen peroxide (OSHA PEL 1 ppm, NIOSH IDLH 75 ppm) are well known to be toxic (otherwise they would not work as disinfectants and sterilants).

The argument has been made that these compounds are normally completely contained within the sterilizers, devices designed to retain these gases/vapors and so leaks are unlikely; but it is unclear how this situation is different from a cold food storage facility using ammonia based refrigeration (OSHA PEL = 50 ppm, NIOSH IDLH = 300 ppm). In both cases, the equipment is designed to keep the the gas inside, but the sterilizer door must be opened on a regular basis to load and unload the items being sterilized. I have never seen a food storage facility or similar establishment using ammonia exposure badges to protect its employees, but it is not unusual to find hospitals that monitor their employees exposure to ethylene oxide or hydrogen peroxide with exposure badges. Continuous monitors are commercially available for ethylene oxide, hydrogen peroxide, ozone, peracetic acid etc, in many cases from several suppliers. The question therefore is why do many hospitals insist on using badges to monitor worker exposure instead of continuous gas monitors.

The National Occupation Research Agenda's (NORA's) Report on workplace safety in healthcare offers an answer: "a key barrier to addressing them [chemical hazards] is the misconception that HCSA [Heathcare sector] work is safer than other work involving exposure to chemical and physical hazards." However a recent Bureau of Labor Statistics report however refutes this assumption.

The same NORA report later comments that "For instance, several authors have pointed out that for many healthcare personnel (HCPs), patient-care issues (i.e., patient health, well-being, and safety) take precedence over personal safety [DeJoy et al. 1995]. There is also a concern that, at least in some settings, a culture and climate of risk acceptance may be the norm; some workers may come to expect that the risk of exposure or injury is simply part of the job."

We often see two versions of this story. The first is an unwillingness to accept the risks posed by the use of the disinfectant and sterilant chemicals in healthcare. This version is easier to understand. The overall risk is a function of the hazard presented by exposure (these are sterilant and disinfectant chemicals after all) and the probability of exposure. If the employer believes, after reasonable investigation, that the risk of exposure is negligible, then the employer can at least argue for not monitoring.

If however, the employer acknowledges that there is risk of exposure, that equipment and engineering controls can sometimes fail and that therefore some kind of monitoring of these gases and vapors is needed then why use an exposure badge, a technology that become obsolete in the 1980s? If an organization uses hazardous chemicals that present a significant risk of exposure, that risk is recognized, and continuous monitors are readily available, then is resorting to badges really meeting the expected standard of care for workplace safety for those employees using these chemicals?

Legal issues aside, the main question for managers who need to decide what type of protection from toxic gases is whether to use continuous monitors, equipment that is widely accepted across many industries to provide good protection for their people or to use exposure badges that have since fallen from main stream use because they don't provide the level of protection needed to keep their workforce safe.

It is dangerous and unfair to paint an industry with too broad a brush and many hospitals have excellent safety records and ensure that their staff are well trained and have modern safety equipment. These facilities should be applauded for setting the standard in healthcare that others would do well to follow.

Monday, June 25, 2012

Questions

The goal of this blog is provide useful information about gas detection and safety, relevant regulations, ChemDAQ's news, industry topics and to be a resource for those who work in these areas. If you have any questions or would like more information about any related topic, please submit a comment below.